The Western Blot technique for identification of specific proteins may have several problems that need to be resolved. One of these is a diffuse or nonspecific band.


Situation: diffuse or nonspecific band

 

Problem

Possible Solution

antibody concentration too high

reduce the concentration of the primary and other antibodies

too much protein loaded onto the gel

reduce the volume of sample loaded onto the gel

signal from chemiluminescent substrate too strong

try one or more of the solutions listed below

 

Possible solutions for a strong signal:

(1) reduce the film exposure time

(2) reduce the concentration of the substrate

(3) shorten the incubation time of the membrane with the substrate

(4) completely remove substrate after the incubation period

(5) decrease the concentration of antibodies, particularly conjugated antibodies